- Bump-test your gas monitor against known gas concentrations before every shift.
- Test the atmosphere at multiple levels before entry using an extension sampling probe.
- Check oxygen is between 19.5% and 23.5% before permitting confined space entry.
- Evacuate immediately if flammable gas readings exceed 10% of the lower explosive limit.
- Maintain continuous atmospheric monitoring throughout the duration of every entry.
- Record all gas test results on the confined space entry permit before authorising entry.
- Ensure your gas monitor is within its calibration date and serviced at required intervals.
- Understand the alarm set points on your monitor and what each alarm level means.
- Re-test the atmosphere after any interruption, disturbance, or change in ventilation.
- Keep the gas monitor on your person, not left at the entry point during entry.
|
- DON'T enter any confined space without testing the atmosphere with a calibrated monitor.
- DON'T test from only one position — sample at high, mid, and low levels in the space.
- DON'T enter if oxygen readings are outside the 19.5% to 23.5% safe range.
- DON'T ignore flammable gas readings — even low levels indicate a hazard that may increase.
- DON'T rely on pre-entry testing alone — conditions change and need continuous monitoring.
- DON'T issue an entry permit without documenting the gas test results on the permit form.
- DON'T use a gas monitor that is overdue for calibration or has failed its bump test.
- DON'T dismiss an alarm as a false positive — evacuate first, then investigate from outside.
- DON'T assume the atmosphere is still safe after a break — re-test before re-entering.
- DON'T leave your gas monitor at the entrance — wear it on your person throughout entry.
See also: Confined Space Awareness | Confined Space Entry Procedures
|