- Isolate process connections and confirm zero pressure before installing any instrument
- Use calibrated test equipment traceable to national standards for all calibration work
- Follow the instrument data sheet and manufacturer instructions for installation and setup
- Maintain cable separation between instrument and power circuits as specified in the design
- Complete loop testing from sensor through to PLC/SCADA before declaring the instrument live
- Wear appropriate PPE for the process area including chemical-resistant gloves and eye protection
- Check ATEX zone classification and use only intrinsically safe instruments and barriers
- Record all calibration values, as-found and as-left readings, and certificate references
- Brief all workers on the process hazards present in the installation area before starting
- Label every instrument and cable clearly with the tag number from the instrument schedule
|
- DON'T install instruments on pressurised pipework without confirmed isolation and venting
- DON'T calibrate instruments with uncertified test equipment — results will be unreliable
- DON'T install non-IS equipment in ATEX-classified hazardous areas without IS barriers
- DON'T route instrument cables alongside power cables without the specified separation distance
- DON'T commission instruments without completing a full loop test from sensor to control system
- DON'T enter confined spaces or process tanks for instrument work without a valid permit
- DON'T force threaded connections on pressure tappings — cross-threading causes dangerous leaks
- DON'T leave instrument covers off or cable glands unsealed exposing electronics to moisture
- DON'T modify instrument ranges or alarm settings without the control engineer's authorisation
- DON'T work in operating process areas without assessing chemical and biological exposure risks
See also: MEICA Safety Awareness | Commissioning Safety
|